![Agilent Technologies G7121A Скачать руководство пользователя страница 177](http://html.mh-extra.com/html/agilent-technologies/g7121a/g7121a_user-manual_2867752177.webp)
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
177
Test Functions and Calibration
8
Raman ASTM Signal-to-Noise Test
Interpretation of the Results
If the test shows low Raman values, check for:
✔
correctly positioned flow cell,
✔
clean flow cell (flush with clean bi-distilled water),
✔
no air bubble(s) (check via fluorescence scan or visual check of
cell/cuvette),
✔
solvent inlet filter (may create air bubbles in flow cell).