Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
109
Optimizing the Detector
5
Example: Optimization for Multiple Compounds
Select a Excitation
wavelength in the
low UV (230...260
nm). This will cover
nearly all
fluorescence in your
sample.
DO NOT select
additional emission
wavelengths (B, C,
D). Doing so will
increase the scan
time and will lower
the performance.
Figure 37
Detector Settings for Emission Scan