![Agilent Technologies G7117C Скачать руководство пользователя страница 209](http://html1.mh-extra.com/html/agilent-technologies/g7117c/g7117c_user-manual_2867751209.webp)
Agilent InfinityLab LC Series Diode Array Detectors User Manual
209
13
LAN Configuration
Initialization Mode Selection
Initialization Mode Selection
The following initialization (init) modes are selectable:
Default IP address for LAN is 192.168.254.11.
DHCP address is the module’s LAN MAC address.
Using Stored
When initialization mode
Using Stored
is selected, the parameters are taken from
the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured previously
by one of the described methods.
Figure 59 Using Stored (Principle)
Table 32
Initialization Mode Switches
SW1
SW2
SW3
SW4
SW5
SW6
Init Mode
0
0
0
0
0
0
Use Default IP Address
0
0
0
0
1
0
Use Stored IP Address
0
0
0
1
0
0
Use DHCP
Note: The setting ‘0’ (down) is essential.
Non-Volatile
RAM
Stored
Parameter
Active
Parameter