Test Functions and Calibration
Wavelength Calibration
Wavelength Recalibration Fails
Probable cause
Suggested actions
1
Absorbing solvent or air bubble in Max-Light
Cartridge Cell.
Repeat calibration with Max-Light Cartridge Test
Cell and compare results.
2
Dirty or contaminated Max-Light Cartridge
Cell.
•
Ensure the Max-Light Cartridge Cell is filled
with water.
•
Recalibrate.
3
Old UV-lamp.
Exchange the UV-lamp.
4
Dirty or contaminated optical components.
Run the Cell Test. If the test fails, flush the flow
cell. See also
“Clean the Max-Light Cartridge
NOTE
If the test fails with Max-Light Cartridge Test Cell and new UV-lamp, the optical
unit must be replaced.