Agilent InfinityLab LC Series Diode Array Detectors User Manual
131
8
Test Functions and Calibration
Intensity Test
Figure 36 Intensity Test – Signals
Test Failed
Intensity Test Evaluation
Probable cause
Suggested actions
1
Absorbing solvent or air bubble in flow cell.
•
Ensure the flow cell is filled with water, and
free from air bubbles.
•
Repeat test with Max-Light Cartridge Test
Cell and compare results.
2
Incorrect calibration
Recalibrate and repeat the test.
3
Dirty or contaminated flow cell.
Run the cell test. If the test fails, flush the flow
cell. See also
“Clean the Max-Light Cartridge
4
Dirty or contaminated optical components.
Please contact your Agilent service representa
-
tive.