Using the Module
Install the Divider Assembly
Install the Divider Assembly
Parts required
p/n
Description
G7116-60006
Divider Assembly MCT
NOTE
The Divider Assembly must be installed if different temperatures are used on the
right and the left heater element to separate these two temperature zones.
1
Push the rear part of the Divider Assembly into position
(1.) and fix it with the screw (2.).
2
Install the valve head if used.
See
3
Install all capillaries and heat exchangers required.
For details, refer to
4
Install the front part of the Divider Assembly by putting
it on top.
5
The front plate should be flush with the top and the bot
-
tom edge of the MCT housing.
1.
2.