Agilent 1290 Infinity II Flexible Pump User Manual
7
Maintenance
Replace the Inlet Weaver
Replace the Inlet Weaver
Parts required
p/n
Description
G4204-81090
1290 Infinity Quaternary Pump Inlet Weaver Assembly
Preparations
•
Switch off pump at the main power switch
•
Open the doors
•
Use an optional solvent shutoff valve or lift up solvent filters inside solvent bottles for avoiding
leakages
•
For easy access to the inlet weaver assembly, remove tubing connections between MCGV and
degasser
1
Open the screw at the bottom of the inlet valve.
2
Open the fitting at the center of the multi-channel
gradient valve (MCGV). Do not open the screw marked
with the black cross. Remove the inlet weaver from the
MCGV.