88
Agilent 1290 Infinity II Flexible Pump User Manual
7
Maintenance
Replace the Pressure Sensor
Replace the Pressure Sensor
When
No or invalid pressure signal
Tools required
p/n
Description
8710-2412
Hex key 2.5 mm, 15 cm long, straight handle
8710-0510
Wrench open 1/4 — 5/16 inch
Screwdriver
Parts required
p/n
Description
G7104-60001
Pressure sensor 1300 bar
Preparations
Turn off pump flow, switch off pump
N O T E
This procedure describes how to replace the pressure sensor.
In case the cable to the sensor shall be replaced as well, please contact your Agilent
service representative.
N O T E
Working on connections to the pressure sensor may slightly change the displayed
pressure. In case of a pressure offset at ambient pressure, a pressure offset calibration may
be run.
1
Remove capillary connections between the pressure
sensor and the High Pressure Filter, and between the
pressure sensor and the outlet filter of the secondary
pump head, respectively.
2
Remove the screws that fix the pressure sensor to the
chassis.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...