![Agilent Technologies G7104A Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/agilent-technologies/g7104a/g7104a_user-manual_2867748074.webp)
74
Agilent 1290 Infinity II Flexible Pump User Manual
6
Error Information
Pump Error Messages
Timeout: Wait for run volume
Error ID: 29181
A target condition (run volume, which is the volume delivered since the
method run start) has been sent to the instrument which should have
been reached within an expected time frame but didn’t. Either the limit,
time frame or the current value of the variable has been modified later
directly or indirectly (for example the flow rate).
Timeout: Wait for Volume
Error ID: 29182
A target condition (volume, which is the delivered flow since the limit has
been set) has been sent to the instrument which should have been reached
within an expected time frame but didn’t. Either the limit, time frame or
the current value of the variable has been modified later directly or
indirectly (for example the flow rate).
Probable cause
Suggested actions
1
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.
Probable cause
Suggested actions
1
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...