Agilent 1290 Infinity II Flexible Pump User Manual
73
Error Information
6
Pump Error Messages
Pump drive stop not found
Error ID: 29207
The pump drive stop has not been found.
•
Parameter: 1 – 2 referring to pump drive
Timeout: Wait for Composition
Error ID: 29180
A target condition (composition) has been sent to the instrument which
should have been reached within an expected time frame but didn’t.
Either the limit, time frame or the current value of the variable has been
modified later directly or indirectly.
Probable cause
Suggested actions
1
Pump drive spindle is defect.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Verify control software, macros, manual
commands.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...