Error Information
6
Pump Error Messages
Writing the multi purpose valve tag failed
Error ID: 29241
Writing the multi purpose valve tag failed.
•
Parameter: None
Unknown multi purpose valve type
Error ID: 29242
The type information of the multi purpose valve is invalid.
•
Parameter: None
Pump drive encoder error
Error ID: 29211
The pump drive encoder has generated no signal.
•
Parameter: 1 – 2 referring to pump drive
Probable cause
Suggested actions
1
Multi purpose valve head tag defect.
Replace multi purpose valve head.
2
Multi purpose valve tag head reader is
defect.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Wrong valve head installed.
Check or replace multi purpose valve head.
2
Valve head has invalid RFID tag content.
Check or replace multi purpose valve head.
Probable cause
Suggested actions
1
Pump drive encoder is defect.
Please contact your Agilent service
representative.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...