36
Agilent 1290 Infinity II Flexible Pump User Manual
3
Using the Pump
Best Practices
Prepare the pump
Purge
Use the Purge function to:
•
fill the pump,
•
exchange a solvent,
•
remove air bubbles in tubes and pump heads.
Condition
Use the Conditioning function:
•
daily when starting the pump,
•
to minimize pressure ripple by dissolving air bubbles in pump heads.
Seal wash
The seal wash function runs continuously and is controlled by the seal
wash sensor. This guarantees a maximum seal life time.
N O T E
Condition your complete system with solvents and composition of your application (for
example 50 %/50 % A/B at 0.5 mL/min.
C A U T I O N
Contaminated seal wash solvent
➔
Do not recycle seal wash solvent to avoid contamination.
➔
Weekly exchange seal wash solvent.
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...