Replace the High Pressure Outlet Filter or Filter Frit
Replace the High Pressure Outlet Filter or Filter Frit
When
For removing blockages and leaks in the high pressure filter assembly. The outlet filter should be
replaced as required depending on the system usage. Other parts are covered by the Agilent
Preventive Maintenance (PM) Service.
Tools required
p/n
Description
8710-0510
Wrench open 1/4 — 5/16 inch
8710-1924
Wrench open 14 mm
Torque wrench
Torque wrench head, 14 mm for torque wrench
Parts required
p/n
Description
G4204-60004
Outlet filter 1290 Infinity Quaternary Pump
5067-5716
Frit for 1290 pump outlet filter 2/pk
1
Remove the capillary from the high pressure outlet filter
to the pressure sensor (1) and from the high pressure
outlet filter to the Multipurpose valve (2).
2
Remove the high pressure outlet filter from the filter
holder .
&#
'#
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...