Maintenance
7
Replace Parts of the Multi Purpose Valve
Replace Parts of the Multi Purpose Valve
1
Use the 9/64 inch hex key for opening the valve head.
2
Replace parts as required.
3
Reassemble the valve head and mount it to the valve drive.
Tools required
p/n
Description
8710-2394
9/64 inch hex key
Parts required
p/n
Description
1535-4045
Bearing ring
5068-0202
Rotor seal, Multi Purpose Valve, 1300 bar
5068-0120
Stator ring
5068-0001
Stator head
1535-4857
Stator screws, 10/Pk
Preparations
•
Remove capillary connections from ports 1, 3 and 6.
•
Remove the clamp with the inline filter.
7ZVg^c\g^c\
GdidghZVa
HiVidgg^c\
HiVidg]ZVY
HiVidghXgZlh
Содержание G7104A
Страница 1: ...Agilent Technologies Agilent 1290 Infinity II Flexible Pump User Manual...
Страница 8: ...8 Agilent 1290 Infinity II Flexible Pump User Manual Contents...
Страница 22: ...22 Agilent 1290 Infinity II Flexible Pump User Manual 1 Introduction Leak and Waste Handling...
Страница 32: ...32 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Magnets Magnets...
Страница 38: ...38 Agilent 1290 Infinity II Flexible Pump User Manual 3 Using the Pump Best Practices...
Страница 160: ...160 Agilent 1290 Infinity II Flexible Pump User Manual 8 Parts and Materials Tools...
Страница 218: ...218 Agilent 1290 Infinity II Flexible Pump User Manual 11 LAN Configuration PC and User Interface Software Setup Setup...