![Agilent Technologies G7104A Скачать руководство пользователя страница 114](http://html1.mh-extra.com/html/agilent-technologies/g7104a/g7104a_user-manual_2867748114.webp)
114
Agilent 1290 Infinity II Flexible Pump User Manual
7
Maintenance
Pump Head Procedures
Disassemble the Secondary Pump Head
C A U T I O N
Damage of pump piston
The pump piston is made of ZrO
2
-based ceramic, which is a very hard and resistant
material, but it is sensitive to shearing forces from the side.
➔
Do not try to remove the pump piston from the rear.
➔
Do not use the piston for removing pump seals.
1
For disassembling the pump head, remove the 4 hex
screws at the rear of the pump head.
2
Remove the front part of the pump head including pump
chamber housing, distance sheet, pump seal and seal
holder, and support ring with wash seal and gasket.
EjbeX]VbWZg
]djh^c\
9^hiVcXZh]ZZi
HZVa]daYZg
HZValVh]\Vh`Zi
Hjeedgig^c\