Operating in Electron Ionization (EI) Mode
3
5977 Series MSD Operation Manual
53
The EI GC/MSD Interface
The GC/MSD interface (
54) is a heated conduit into the MSD
for the capillary column. It is bolted onto the right side of the analyzer
chamber, with an O-ring seal. It has a protective cover which should be left in
place.
One end of the GC/MSD interface passes through the side of the gas
chromatograph and extends into the GC oven. This end is threaded to allow
connection of the column with a nut and ferrule. The other end of the
interface fits into the ion source. The last 1 to 2 mm of the capillary column
extend past the end of the guide tube and into the ionization chamber.
The GC/MSD interface is heated by an electric cartridge heater. Normally, the
heater is powered and controlled by Thermal Aux #2 heated zone of the GC.
The interface temperature can be set from MassHunter Data Acquisition or
from the gas chromatograph. A sensor (thermocouple) in the interface
monitors the temperature.
The GC/MSD interface should be operated in the 250
to 350
C range. Subject
to that restriction, the interface temperature should be slightly higher than the
maximum GC oven temperature, but
never
higher than the maximum column
temperature.
The EI GC/MSD interface can only be used with an EI ion source. There are
two types of tip seals that may be used with EI ion sources. The standard/inert
EI ion sources do not need a tip seal. The extractor ion source on an EI only
system uses a tip seal (G3870-20542). The extractor ion source and the CI ion
source on an EI/CI system uses the CI tip seal (G1999-60412).
See also
“To Install a Capillary Column in the GC/MS Interface”
WA R N I N G
The GC/MSD interface operates at high temperatures. If you touch it when it is hot,
it will burn you.
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...