CI Maintenance
6
5977 Series MSD Operation Manual
163
3
Clean the parts as described in
C A U T I O N
If the insulators are dirty, clean them with a cotton swab dampened with reagent-grade
methanol. If that does not clean the insulators, replace them. Do not abrasively or
ultrasonically clean the insulators.
Figure 30
CI ion source parts to be cleaned
Entrance lens
Ion focus lens
Drawout
cylinder
Drawout lens
Source body
Repeller