General Maintenance
5
5977 Series MSD Operation Manual
151
To Replace the Electron Multiplier Horn
Materials needed
•
Electron multiplier horn (G3170-80103)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
Procedure
1
Vent the MSD. See
2
Open the analyzer chamber. See
“To Open the Analyzer Chamber”
3
Open the retaining clip. Lift the arm of the clip up and then swing the clip
away from the electron multiplier horn.
Figure 28
Electron multiplier horn
Electron multiplier horn
Retaining clip