116
5977 Series MSD Operation Manual
4
Operating in Chemical Ionization (CI) Mode
To Verify NCI Performance
This procedure is for EI/PCI/NCI MSDs
only
.
Materials needed
•
Octafluoronaphthalene (OFN), 100 fg/µL (5188-5347)
Procedure
1
Verify that the MSD performs correctly in EI mode.
2
Load the
NCICH4.U
tune file, and accept the temperature setpoints.
3
Select
Gas A
and set flow
to 40%.
4
In Tune and Vacuum Control view, run CI Autotune. See
NCI Autotune (Methane Reagent Gas)”
Note that there are no criteria for a “passing” Autotune in CI. If the
Autotune completes, it passes.
5
Run the NCI sensitivity method: OFN_NCI.M using 2 µL of 100 fg/µL OFN.
6
Verify that the system conforms to the published sensitivity specification.
Please see the Agilent Web site at
for
specifications.
C A U T I O N
Always verify MSD performance in EI before switching to CI operation. See
71. Always set up the CI MSD in PCI first, even if you are
going to run NCI.
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...