![Agilent Technologies G7035A Скачать руководство пользователя страница 115](http://html1.mh-extra.com/html/agilent-technologies/g7035a/g7035a_operation-manual_2868127115.webp)
Operating in Chemical Ionization (CI) Mode
4
5977 Series MSD Operation Manual
115
To Verify PCI Performance
Materials needed
•
Benzophenone, 100 pg/
L (8500-5440)
Procedure
1
Verify that the MSD performs correctly in E1 mode.
2
Verify that the
PCICH4.U
tune file is loaded.
3
Select
Gas A
and set flow to 20%.
4
In
Tune and Vacuum Control
view, perform CI setup. See
CI Ion Source to the Extractor EI Ion Source”
5
Run CI Autotune. See
6
Run the PCI sensitivity method
BENZ_PCI.M
using 1 µL of 100 pg/µL
benzophenone.
7
Verify that the system conforms to the published sensitivity specification.
Please see the Agilent Web site at
for
specifications.
C A U T I O N
Always verify MSD performance in EI before switching to CI operation. See
71. Always set up the CI MSD in PCI first, even if you are
going to run NCI.
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...