background image

 

 

 

Purpose 

The Brass Burner has two ceramic combustion tubes which require periodic 
replacement:  an inner and an outer tube. 

Scope 

This procedure applies to ceramic tube replacement in the Agilent Instruments 
ABI 00003 Brass Burner. 

Tools 

 

2 - 1/16” Open End Wrenches 

 

2 - 3/8” Open End Wrenches 

 

2 - 7/16” Open End Wrenches 

Procedure 

1.

 

Turn off the Interface Controller and let the system cool down 

under 

vacuum

The Burner is extremely hot when running.  Do not touch the burner until it has 
cooled down. 

2.

 

Turn off the vacuum pump. 

3.

 

Remove the hydrogen line by disconnecting the Hydrogen Inlet Fitting at 
the top of the Burner.  Use a second wrench to hold the Adapter Tee to 
avoid excess torque on the ceramic tubes. 

4.

 

Disconnect the black PFA transfer line from the Tee of the Burner 
Interface. 

5.

 

Unscrew the Tee fitting and carefully pull the inner Tube/Tee assembly 
from the Burner.  

If the existing Tube is broken then all fragments must 

be removed from the burner.

 

Note 

Ceramic Probe Replacement 
Instructions 

Accessory G6600-60034 

Содержание G6600

Страница 1: ...r is extremely hot when running Do not touch the burner until it has cooled down 2 Turn off the vacuum pump 3 Remove the hydrogen line by disconnecting the Hydrogen Inlet Fitting at the top of the Burner Use a second wrench to hold the Adapter Tee to avoid excess torque on the ceramic tubes 4 Disconnect the black PFA transfer line from the Tee of the Burner Interface 5 Unscrew the Tee fitting and ...

Страница 2: ...er 9 Remove the four horizontal socket head cap screws holding the Burner Case 10 Carefully lift the Burner Case up off of the base and set it aside 11 Remove the white insulating disk and the fiber insulating toroid from the Burner base 12 Remove the 1 8 ceramic tube by loosening the nut in the Burner base Discard the used ceramic tube and ferrule Remove any debris from the inside of the connecti...

Страница 3: ...loosen under heat stress but excessive tightening can crack the ceramic tube 14 Reinstall the fiber insulating toroid and the white insulating disk Carefully slide the Burner Cover over the large ceramic tube It is often useful to sight down the upper fitting to ensure that the ceramic tube is properly aligned and will not break Failure to install the insulating disk may lead to leaks in the Burne...

Страница 4: ...nto the Burner tightening finger tight 19 Check the flow of air and hydrogen following the Detector manual instructions 20 Reconnect the Heater Cable Thermocouple and all the tubing Pressure test the system following the instructions in the Detector manual The pressure may change slightly due to variation in tube dimensions 21 Turn on the Interface Controller and allow the Burner temperature to st...

Страница 5: ......

Страница 6: ...Copyright 2007 Agilent Technologies Inc 2850 Centerville Road Wilmington DE 19808 1610 Printed in the USA Feb 2007 ...

Отзывы: