background image

Agilent G6011B

Quiet Cover MS

User Manual

Содержание G6011B

Страница 1: ...Agilent G6011B Quiet Cover MS User Manual...

Страница 2: ...on flict with these terms the warranty terms in the separate agreement shall control Technology Licenses The hardware and or software described in this document are furnished under a license and may b...

Страница 3: ...reparation 11 2 Installation Installing the Oil Tray 13 Accommodating a KF40 Foreline Pump Intake Flange 14 Installing the Rough Pump Inside the Quiet Cover MS 15 Connecting the Power Supply 18 Verify...

Страница 4: ...pumps Your rough pump fits in the Quiet Cover MS to significantly reduce noise while an internal fan ensures that your pump will not overheat There is an easy to use kickstand and a full size oil tra...

Страница 5: ...r a quality system registered to ISO 9001 Information The Agilent Technologies Quiet Cover MS meets the following IEC classifications Category II Pollution Degree 2 Class III Quiet Cover MS has been d...

Страница 6: ...violates safety standards of design and the intended use of the instrument Agilent Technologies assumes no liability for the customer s failure to comply with these requirements Table 1 lists the saf...

Страница 7: ...ure to perform a two person lift may result in personal injury WARNING Never run the pump without also running Quiet Cover MS CAUTION While running Quiet Cover MS Do not block air flow to the fan Do n...

Страница 8: ...evision are on separate electrical circuits Make sure that all peripheral devices are also certified Make sure that appropriate cables are used to connect the device to peripheral equipment Consult yo...

Страница 9: ...21 for a view of the front panel control Table 2 G6011B Quiet Cover MS contents list Part Quantity Part number Quiet Cover MS 1 G6011B Oil tray 1 G6011 40100 Power supply 1 5188 1164 KF40 Trim ring 1...

Страница 10: ...1 Parts Identification Parts Identification Quiet Cover MS User Manual 10 Figure 2 Quiet Cover MS rear view power supply not shown Kickstand Power connection Fan Intake...

Страница 11: ...ion Physical dimensions Weight 11 3 kg 25 lbs Dimensions Kickstand up 68 25 44 cm 27 13 5 17 25 inches L W H Power supply AC input voltage 100 to 240 VAC 10 DC output voltage 24 V Output load max 1 25...

Страница 12: ...Tray 13 Accommodating a KF40 Foreline Pump Intake Flange 14 Installing the Rough Pump Inside the Quiet Cover MS 15 Connecting the Power Supply 18 Verifying the Installation 19 This chapter describes...

Страница 13: ...y You need to install it prior to using the Quiet Cover MS 1 Remove any protective film from the tray 2 Slide the tray into the bottom of the Quiet Cover MS with the deep part of the tray facing the f...

Страница 14: ...ng the G6011B Quiet Cover on an MS 40 vacuum pump with a KF25 Foreline Pump intake flange skip this procedure If you are installing the G6011B Quiet Cover on an MS 40 vacuum pump with a KF40 Foreline...

Страница 15: ...modate an MS 40 vacuum pump If this revision of pump is being installed no changes are required and the installer can proceed to the next procedure on page 16 If an MS 40S pump is being installed comp...

Страница 16: ...6 Locking casters top access panels and main cover 2 To remove the main cover assembly a Open the front access door b Lift the main cover vertically to detach it from the guide pins WARNING A two per...

Страница 17: ...able If the support feet do not align with the mounting cup and mounting tabs the rough pump orientation is incorrect Pump should be slightly touching both the motor air duct and the electrical panel...

Страница 18: ...perating location unlock the locking casters carefully push the Quiet Cover MS to its operating location and lock the front locking casters WARNING Do not attempt to lift the Quiet Cover MS with the r...

Страница 19: ...e Quiet Cover MS 3 Turn the Power Switch to the off position 4 Plug the G6011B Quiet Cover into an external power outlet If the instrument is receiving power the Power Switch Figure 1 should light up...

Страница 20: ...Pump Oil 22 Using the Kickstand 24 To operate Quiet Cover MS you will need to Understand the controls and indicators on the front panel Know how to handle temperature alarms and signals Know how to d...

Страница 21: ...atuses are normal operating modes If the air flow to the Quiet Cover MS is restricted or if the room temperature is above 35 C 95 F the red status indicator on the front panel lights and the temperatu...

Страница 22: ...S 4 Unplug the power supply from the Quiet Cover MS 5 Lock the locking casters to secure the Quiet Cover MS 6 Open the front access door of the Quiet Cover MS 7 Tilt the Quiet Cover MS by using the ki...

Страница 23: ...ironmental guidelines 12 Return the Quiet Cover MS to its untilted position See Using the Kickstand on page 24 13 Reinsert the oil tray 14 Remove the oil fill plug Figure 12 and refill the pump Figure...

Страница 24: ...Quiet Cover MS is on a level surface 2 Lock the two locking casters located on the front of the Quiet Cover MS Figure 14 3 Release the kickstand from the retaining clip Figure 15 Figure 14 Locking cas...

Страница 25: ...the floor when Quiet Cover MS is tilted To return Quiet Cover MS to the untilted position 1 Grasp the back of the Quiet Cover MS and gently push the Quiet Cover MS forward off the kickstand To avoid...

Страница 26: ...ump oil does not drain completely Be sure the kickstand is down with the bar sitting flat on the floor If so refer to the rough pump documentation for more information Green LED does not light Check t...

Страница 27: ...www agilent com Agilent Technologies Inc 2020 Second edition April 2020 G6011 90110 G6011 90110...

Отзывы: