82
1260 SFC Binary Pump
6
Error Information
Module Error Messages
Temperature Out of Range
Error ID: 2517
Temperature Out of Range 0:
Pump channel A
Temperature Out of Range 1:
Pump channel B
One of the temperature sensor readings in the motor-drive circuit are out of
range.
The values supplied to the ADC by the hybrid sensors must be between 0.5 V
and 4.3 V. If the values are outside this range, the error message is generated.
Temperature Limit Exceeded
Error ID: 2517
Temperature Limit Exceeded 0:
Pump channel A
Temperature Limit Exceeded 1:
Pump channel B
The temperature of one of the motor-drive circuits is too high.
The processor continually monitors the temperature of the drive circuits on
the main board. If excessive current is being drawn for long periods, the
temperature of the circuits increases. If the temperature exceeds the upper
limit, the error message is generated.
Probable cause
Suggested actions
1
Defective main board.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
High friction (partial mechanical blockage)
in the pump drive assembly.
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
2
Partial blockage of the flowpath in front of
the damper.
Ensure the outlet valve is not blocked.
3
Defective pump drive assembly.
Please contact your Agilent service
representative.
4
Defective main board.
Please contact your Agilent service
representative.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...