1260 SFC Binary Pump
107
Maintenance
8
Maintenance of a Pump Head without Seal Wash
Maintenance of a Pump Head without Seal Wash
When
In case of maintenance or pump head internal leaks
Tools required
p/n
Description
8710-0510
Wrench open 1/4 — 5/16 inch
8710-2411
Hex key 3 mm12 cm long
8710-2392
Hex key 4 mm15 cm long T-handle
01018-23702 Insert
tool
Parts required
#
p/n
Description
1
5063-6589
Piston seal PTFE, carbon filled, black (pack of 2), default
OR
1
0905-1420
PE seals (pack of 2)
1
5067-4695
Sapphire piston
Preparations
•
Switch off the pump at the main power switch.
•
Open the doors of the module.
•
Use an optional solvent shutoff valve or lift up solvent filters for avoiding leakages.
•
Remove the Pump Head Assembly.
N O T E
Both pump head assemblies use the same internal components. In addition, pump head A
is fitted with the purge valve. The following procedure describes the removal and
disassembly of pump head A (left). For pump head B (right) proceed in the same way and
skip steps that deal with the purge valve.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...