Maintenance
Replace the O-Ring on the Purge Valve
Replace the O-Ring on the Purge Valve
When
If the original o-ring is damaged and needs to be replaced
Tools required
p/n
Description
8710-0510
Wrench open 1/4 — 5/16 inch
8710-1924
Wrench open 14 mm
Pair of tweezers
OR
Toothpick
Parts required
#
p/n
Description
1
5067-6595
1260 PV O-ring FKM 5/pack
1
01018-22707
PTFE frits (pack of 5) (OPTIONAL)
1
5067-4728
Seal cap (OPTIONAL)
Preparations
•
Switch off pump at the main power switch.
•
Open the doors of the module.
•
Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.
•
Remove the purge valve from the pump head.
1
Disassemble the purge valve.
2
Remove the old o-ring from the purge valve.
3
Clean the purge valve parts.