Installation and Operation Guide
1
Introduction
The column flow mixes with the makeup flow in the splitter. This mixture then
flows through lengths of uncoated, deactivated, fused-silica tubing to each
detector. These tubes act as flow restrictors. While the flows through the
restrictors change with oven temperature, the
ratio
of the flows at any
temperature is constant.
Details
The G3183B kit addresses several limitations of previous approaches to
splitting column effluent between three detectors:
Metal ferrules
The splitter uses metal column ferrules, which eliminate air leakage into the
sample stream. Unlike polyimide, metal ferrules do not loosen upon thermal
cycling of the oven. They also do not outgas contaminants or shed particles
(like graphite) that can result in chromatographic problems.
Microfluidic plate
The splitting hardware is based on microfluidic plate technology. This allows
very low dead volume connections between the column end and the three
detector restrictor tubes. The thin metal plate has fast thermal response and is
mounted solidly on the oven wall for ease of use. The interior plate surfaces
are deactivated to prevent adsorption of active compounds.
Constant pressure operation
The splitter uses a source of makeup gas supplied by electronic pneumatics
control (EPC). This maintains the splitter at a known and constant pressure.
Constant pressure allows easier splitting to vacuum detectors like the MSD. It
simplifies choice of splitter parameters, allowing all aspects of the
chromatographic setup to be calculated. Constant pressure makeup allows the
column to be run in constant flow mode while still maintaining a constant
split ratio between three detectors of different operating pressures such as the
FPD and the MSD. Because the EPC pressure can be time programmed, useful
operations like backflushing unwanted heavy materials from the column and
changing columns in MSD systems without venting are possible.
Other uses
Other useful configurations can be set up as described in
.
Содержание G3183B
Страница 1: ...Agilent Technologies Agilent G3183B Three Way Splitter Kit Installation and Operation Guide...
Страница 4: ...4 Installation and Operation Guide...
Страница 38: ...38 Installation and Operation Guide 3 Splitter Configurations...
Страница 55: ......
Страница 56: ...Agilent Technologies Agilent Technologies Inc Printed in USA April 2006 G3183 90120...