To adjust or replace the nebulizer needle
8
Nebulizer Adjustment Fixture User Guide
3
Install the nebulizer in the
4
Loosen the locknut next to the zero-dead-volume (ZDV) union.
5
Remove the union from the nebulizer
6
Loosen the fastener (locknut, 1.5-mm hex screw, or T6 Torx screw) that
secures the
in place.
Figure 4
Nebulizer needle holders
Nebulizer with locknut (original needle holder).
Nebulizer with 1.5-mm hex screw.