Entering a Method
User Information
97
Setting Event Times
Use the EVENT TIMES screen to enter values in minutes for the
timed events which make up the vial sampling sequence. The
following events are included:
•
GC Cycle Time
•
Vial Equilibration Time
•
Pressurization Time
•
Loop Fill Time
•
Loop Equilibration Time
•
Inject Time
These times may be checked on the display and changed while
the Headspace Sampler is running samples. Changes are
applied to the next vial analyzed. You cannot change
parameters for a vial currently being sampled. The keys
involved are locked.
Navigate to the EVENT TIMES screen
Press the
Active Method
key. Select
Event Times
from the
Active Method menu. The sampler displays the following:
EVENT TIMES
SET.
GC Cycle Time (min) 25.0
Vial Eq. Time (min)
15.0
Pressuriz. Time (min) 0.20
New Setpoint:
27.0
EVENT TIMES
SET.
Loop Fill Time (min)
0.15
Loop Eq. Time (min) 0.05
Inject Time (min)
0.30
New Setpoint:
0.08
medium_standard.book Page 97 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...