User Information
Maintaining the Instrument
Leak Test
A leak test checks for the following:
•
Restrictions in the flow path associated with filling the loop
•
Leaks in the carrier flow path
•
Leaks in the vial pressurization flow path
•
Leaks across the 6-port sampling valve
The most common leaks are at the connections of the transfer
line, external plumbing, and sample probe union.
Required tools
•
Beaker of water
•
20-mL capped vial
•
Leak test kit G1888-60701
•
7 × 8-mm wrenches
•
This test can be facilitated using the Advanced Function 5
diagnostic tool located on the CD ROM.
Setup
1
Attach the 1/8-inch od tubing to the vent with a Swagelok
connection. Finger tighten the connection.
2
Place the free end of the tube in a beaker of water.
3
If needed, install a shut-off valve between the gas source and
the bulkhead fitting for carrier and vial pressurization gas.
Install the valves on the back of the Headspace Sampler
between the bulkhead fittings and the gas source.
4
For EPC setups, add a T-fitting connecting the shutoff valves
on the carrier gas and vial pressurization gas inlets.
5
Change the pressure units to PSI. This setting has better
resolution than kPa or Bar. See
6
Set the source pressure to 30 PSI.
medium_standard.book Page 148 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...