54
1100 Series WS MWS Reference Manual
2
Optimizing Performance
Choice of Seat Capillary
Different models of seat capillaries are available for the well-plate sampler and
the micro well-plate sampler:
For the Well-plate
Sampler
The needle seat assembly includes the needle seat and the seat capillary. The
part number for this assembly is: G1367-87101.
For the Micro
Well-plate
Sampler
The needle seat assembly is made up of two parts:
•
needle seat: G1377-87101.
•
seat capillary
choice of: G1375-87317 (100 µm 150 mm)
G1375-87300 (50 µm 150 mm)
G1375-87317 (100 µm)
is the capillary preinstalled in micro well-plate and
thermostatted micro well-plate samplers upon delivery.
This capillary is recommended for applications with a 0.3 mm column or
higher. It provides less plugging of the capillary in general and especially with
biological samples. For small K’ this capillary can provide a higher peak width
for isocratic analysis.
G1375-87300 (50 µm) is available as a spare part and is recommended for
applications with a 0.3 mm column or smaller. This capillary gives full
chromatographic performance.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......