Stay-Resident Settings
Firmware update procedures may require this mode in case of firmware
loading errors.
Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident part, that is, it is not operable as
a detector. It only uses basic functions of the operating system for example, for
communication.
To return to normal operation, set switches back to your GPIB or RS-232C
configuration settings.
Table 76
Stay Resident Settings
Mode Select
1
2
3
4
5
6
7
8
TEST/BOOT
1
1
0
0
1
0
0
0
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......