104
7
Maintenance
Simple Repairs
Interface Board
1
Switch OFF the autosampler at the main power switch.
2
Disconnect cables from the interface board connectors.
3
Loosen the screws. Slide out the interface board from the autosampler.
4
Install the interface board. Secure the screws.
5
Reconnect the cables to the board connectors
Figure 22
Exchanging the Interface Board
When
At installation or when defective.
Tools required
•
Flat-head screwdriver.
Parts required
#
Description
1
Interface board.
C A U T I O N
Electronic boards are static sensitive and should be handled with care so as not to
damage them. Touching electronic boards and components can cause electrostatic
discharge (ESD).
ESD can damage electronic boards and components.
➔
Be sure to hold the board by the edges and do not touch the electrical components.
Always use an ESD protection (for example, an ESD wrist strap) when handling
electronic boards and components.
HXgZlh
Содержание G1329A
Страница 1: ...Agilent 1260 Infinity Preparative Autosampler User Manual...
Страница 6: ...6 Contents...
Страница 22: ...22 1 Introduction to the Autosampler Agilent 1200 Infinity Interfaces...
Страница 32: ...32 2 Site Requirements and Specifications Performance Specifications...
Страница 52: ...52 3 Installing the Autosampler Transporting the Autosampler...
Страница 74: ...74 6 Troubleshooting and Diagnostics Overview of the Sampler s Indicators and Test Functions...
Страница 106: ...106 8 Parts and Materials for Maintenance Main Assemblies Main Assemblies Figure 23 Autosampler Main Assemblies...
Страница 109: ...109 Parts and Materials for Maintenance 8 Analytical Head Assembly Figure 24 Analytical Head Assembly cdi h dlc...
Страница 136: ...136 9 Identifying Cables RS 232 Cables...
Страница 149: ...149 Index...