102
7
Maintenance
Simple Repairs
Gripper Arm
When
Defective gripper arm
Tools required
•
Straightened paper clip.
Parts required
#
Part number
Description
1
G1313-60010
Gripper assembly
Preparations
•
Select “Start” in the maintenance function “ChangeGripper”.
•
Turn off the power to the autosampler.
•
Remove the front cover.
1
Identify the slit below the gripper motor and the gripper
arm release button.
2
Rotate the arm approximately 2.5cm (1 inch) to the left
and insert the straightened paper clip into the slit.
Wjiidc
ha^i