86
1200 Series Handheld Control Module Users’s Guide
6
Using the Manual Injection Valve
Partially Filling the Sample Loop
Partially Filling the Sample Loop
Use this method when only small quantities of sample are available. In this
method the syringe determines the injection volume.
1
With valve in INJECT position, use needle port cleaner to flush needle port
with about 1 ml of mobile phase (this will reduce residual contamination
from previous injection).
2
Move valve to LOAD position.
3
Fill syringe with required volume of sample (not more than half of loop
volume).
4
Insert needle of syringe into needle port until needle touches stator face.
Do not press too hard.
5
Slowly inject sample.
6
Leave syringe in position and move valve to INJECT position.
7
Remove syringe.
How Much Sample Is Actually Injected?
When the sample loop is completely filled, the amount of sample injected is
equal to the volume of the sample loop plus the volume of the valve passages
(1 in rotor and 2 in stator). This means that the actual amount of sample
injected will be different to the nominal value designated to your sample loop.
However, since both standards and samples are analyzed using the same loop,
you rarely need to know the absolute volume of the loop.
If you do need to know the actual volume of a sample loop, we recommend you
calibrate it fitted to the valve, so that you also take the valve passages into
account.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......