80
1200 Series Handheld Control Module Users’s Guide
5
Using the Autosampler
Aligning Transport Arm/Gripper
Aligning Transport Arm/Gripper
1
Press the
Tests
button [
F3
] in the
System
screen.
2
Choose Autosampler from the list.
3
Press the
Align
button [
F1
].
4
Choose Transport.
5
Enter the desired values in the corresponding fields.
The field X Correction sets the default value for the axis parallel to the front
panel (left-right motion).
Theta Correction describes the default angle of the transport arm towards
the front-back axis of the module (rotation).
For detailed information on the axes consult your Autosampler reference
manual.
6
Select the
Done
button [
F6
] to accept settings.
The new values will be valid after performing the next hardware initialization
or reset.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......