62
1200 Series Handheld Control Module Users’s Guide
2
Working with the Control Module
Coexecution with Agilent ChemStation
Restrictions
•
If a parameter window is open for parameter entry on the Agilent
ChemStation, this specific entry field is disabled on the control module.
•
If an analysis is running with the control module, the Agilent ChemStation
should not be turned on or rebooted.
•
If the Agilent ChemStation starts an analysis, the control module is the
slave/monitor system.
•
Parameter changes to a method will be identified on the other user
interface as modification.
•
The control module and the Agilent ChemStation have a different method
handling (Agilent ChemStation method can have more information than the
method on the control module, e.g. additional DAD parameters that are only
accessible from the Agilent ChemStation). To have a method available on
both controllers proceed as follows:
If the method is on the Agilent ChemStation and should be saved on the
control module or PC card, load the method on Agilent ChemStation and
then save the method on the control module (or PC card) with Method —
Save As.
If the method is on the control module or PC card and should be saved on
the Agilent ChemStation, first load method DEF_LC.M on the Agilent
ChemStation (to have no additional parameter in the format) and then load
the required method on control module. Then save the method on the
Agilent ChemStation with the same name.
WA R N I N G
A method that is available on the control module as protected method can be
modified by the Agilent ChemStation and then be saved on the control module
without any warning.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......