Time Programming
Time Programming
To time-program selected settings during the analysis you can create a
timetable. Using the Timetable screen, you can create a time based program
that will automatically control the pump, detector, column compartment and
external contacts.
In some cases the settings will change instantaneously from the initial value to
the value specified after a certain time in the timetable (e.g. wavelength). In
other cases (solvent composition) these changes take place dynamically,
approaching the set value in a step wise and linear manner.
The timetable is accessed by selecting the
Timetable
button [
F2
] in the
Analysis
screen.
A timetable line can be inserted by pressing the
Insert
button [
F7
] and consists
of the following:
•
Time
Set the time span between the instant of injection and the desired
parameter change.
•
Module
Choose the module that controls the parameter you want to change.
•
Setting
Select the parameter to be changed.
•
Value
Enter the desired parameter value.
You can edit an existing timetable line by pressing the
Enter
key. Use the
Delete
button [
F6
] and make a choice from the pop-up menu to delete either
the selected line or the whole timetable.
You can copy and paste timetable lines by selecting the respective choices
available from the context menu (
m
key).
N O T E
The timetable becomes part of the current method when the method is saved.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......