1200 Series Handheld Control Module Users’s Guide
145
Running Multiple-Vial Analyses
11
Synchronizing Analyses with External Devices
Send Single Start Request
After you start the analysis with the control module it will generate a single
start request on the APG remote lines. This allows to trigger the external
device which will start each injection by sending a start signal. The vial range
or sequence will be started by the control module, but for every new injection
the external device must give the start command.
In this mode the control module will simply track the progress of the vial
range or sequence. It will indicate the proper sample, change sequence lines,
load methods or other sequence related actions for each run.
Figure 18
Send single external start request
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......