Running Multiple-Vial Analyses
Synchronizing Analyses with External Devices
Synchronizing Analyses with External Devices
With an APG remote connector the system can be connected to external
devices in order to synchronize the analyses. This is necessary when an
external device needs some time in order to get ready for a new analysis and
when transmitting of a start request is required (see
on page 57 for further information on interfaces).
When executing an analysis by pressing the
Start
button [
F8
] you will see a
window where sequence modes can be chosen.
In any case, do all the analysis preparation using the control module.
Standard
In the standard mode the analysis is under the command of the control
module. The control module will issue a Start Request command to the
autosampler as soon as all modules are ready for the next analysis. The
autosampler issues the Start command at the point of injection. With an
Agilent 1200 Autosampler integrated in the system and no external devices
this is the normal operation mode.
N O T E
A “Start” command is used to start the analytical run from the point of injection and is
usually issued by the autosampler.
A “Start Request” command causes the autosampler to take the next vial and place it under
the injection needle (also see
on page 57 and
The “Start button” [F8] on the control module is used to start a vial range or sequence
analysis.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......