1200 Series Handheld Control Module Users’s Guide
137
Running Multiple-Vial Analyses
11
Single-Level Calibration Sequences
Single-Level Calibration Sequences
The following procedure describes how to set up a calibration sequence for an
analysis which uses single-level calibration.
There is one calibration standard (C) and 9 samples (S).
The analysis requires that:
•
each sample is analyzed in duplicate,
•
the calibration standard is analyzed once before the samples and
re-analyzed once after every 2 samples,
C
S S
C
S S
C
S S
C
S S
C
S
C
•
the calibration standard is located in position 90 and the 9 sample vials are
in positions 1 to 9 of the autosampler tray, and
•
the method called METH1 is used for the samples and standards.
The example assumes that the sequence table is empty. If the sequence table is
not empty, use the All Lines item, accessed by pressing the
Delete
button [
F6
].
1
Select the
Sequence
button [
F4
] in the Analysis screen.
2
For sequence line 1, select the
Insert
button [
F7
] and
enter
:
3
Select the
Calibration
button [
F1
] to display the
Calibration Settings
screen for
sequence line 1 and choose
Edit
.
Vial Range
1 to 9
#Inj.
2
Inj. Volume
Default
Method
METH1
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......