1200 Series Handheld Control Module Users’s Guide
135
Running Multiple-Vial Analyses
11
Analyzing Multiple Vials Using Different Methods
Analyzing Multiple Vials Using Different Methods
This section describes how to set up a 50-vial analysis using three methods
which you have previously created called e.g. METH1, METH2 and METH3. For
example: METH1 and METH2 have the same analytical settings but differ in
the injection volume and stoptime values. METH3 uses a different temperature
and requires a wait time of 30 minutes for the LC system to equilibrate.
•
The first 20 vials are analyzed using METH1 with one injection per vial,
•
the next 20 vials are analyzed using METH2 with two injections per vial,
•
the last 10 vials are analyzed using METH3 with three injections per vial.
The vials are located in positions 1 to 50 of the autosampler tray.
The example assumes that the sequence table is empty. If the sequence table is
not empty, use the All Lines item, accessed by pressing the
Delete
button [
F6
].
1
Select the
Sequence
[
F4
] in the
Analysis
screen.
2
For sequence line 1, select the
Insert
button [
F7
] and
enter
:
3
VSelect the
Enter
button to accept entries.
4
For sequence line 2, enter:
N O T E
This way of combining methods is called
Sequence
.
Vial Range
1 to 20
#Inj.
1
Inj. Volume
Default
Method
METH1
Vial Range
21 to 40
#Inj.
2
Inj. Volume
Default
Method
METH2
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......