134
1200 Series Handheld Control Module Users’s Guide
11
Running Multiple-Vial Analyses
Analyzing Multiple Vials Using the Same Method
Analyzing Multiple Vials Using the Same Method
This section describes how to set up a 25-vial analysis with one injection from
each vial. You will use a previously created method. The samples are located in
positions 1 to 25 of the autosampler tray. For details
1
Enter the
Analysis
screen.
2
Set the Vial Range from 1 to 25.
3
Set the number of injections to 1.
4
Select the method you want to use with the scroll buttons (
).
5
Select the
Start
button [
F8
].
6
Select OK to confirm the vial range and injection number and press
Enter
to start the analysis.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......