Entering Settings
If you have a quaternary pump set %B to 65, %C and %D to OFF.
If you have a binary pump set %B to 65.
13
Select the
Pressure
button.
14
Set the upper pressure limit to 400 bar.
15
Select the
Done
button.
16
Select the
Timetable
button.
17
Select the
Insert
button.
18
Enter a value of 2 minutes and select Composition as the setting.
If you have a quaternary pump, set the %B composition to 65, %C and %D to
0.00.
If you have a binary pump, set the %B composition to 65.
19
Select the
Enter
button to accept entries (Action Key
F7
).
20
Enter the following information for another two lines.
time 4 : %B = 95
time 5 : %B = 65
21
Press the
Done
button and verify your timetable entries.
22
Press the
Esc
key until you are in the
Analysis
screen.
N O T E
The channels of a pumping system are named A, B, C and D (depending on the pump type).
%A is automatically calculated by 100% - (%B + %C + %D). If no values for %B, %C and %D
are entered, %A is always 100%.
N O T E
Use Enter to store timetable lines.
Use Done when all lines have been entered.
By pressing the Esc key you return to the previous screen.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......