1200 Series Handheld Control Module Users’s Guide
125
Running a Gradient Analysis
10
Preparing the LC System
Preparing the LC System
1
Fill one solvent bottle with bidistilled water (channel A) and the other with
acetonitrile (channel B).
2
Turn on the detector lamp and pump using the
On/Off
button [
F7
] in the
Analysis
screen.
3
For the quaternary pump, turn on the degasser by pressing the line-power
switch.
4
Purge the pump. For more information see
5
Allow the detector at least 15 minutes to provide a stable baseline.
6
Fill the contents of a Agilent Technologies isocratic standard sample
ampoule into a vial and seal the vial with a cap. Place the vial in position 1
of the autosampler tray.
7
Pump the water/acetonitrile (35/65 %) mobile phase through the column for
10 minutes at a flow rate of 2 ml/min.
Содержание G1323A
Страница 1: ...A Agilent 1200 Series Handheld Control Module User s Guide ...
Страница 14: ...14 1200 Series Handheld Control Module Users s Guide Contents ...
Страница 74: ...74 1200 Series Handheld Control Module Users s Guide 4 Using the Degasser Changing Solvents ...
Страница 100: ...100 1200 Series Handheld Control Module Users s Guide 7 Using the Detectors Configuring the Analog Channel Output ...
Страница 112: ...112 1200 Series Handheld Control Module Users s Guide 8 Using the Column Compartment Resetting the Column Compartment ...
Страница 122: ...122 1200 Series Handheld Control Module Users s Guide 9 Running an Isocratic Analysis Starting the Analysis ...
Страница 132: ...132 1200 Series Handheld Control Module Users s Guide 10 Running a Gradient Analysis Starting the Analysis ...
Страница 154: ...154 1200 Series Handheld Control Module Users s Guide 12 Running an Injector Program Saving the Method ...
Страница 162: ...162 1200 Series Handheld Control Module Users s Guide A Appendix Agilent Technologies on Internet ...
Страница 169: ......