![Agilent Technologies G1314B Скачать руководство пользователя страница 96](http://html1.mh-extra.com/html/agilent-technologies/g1314b/g1314b_user-manual_2868497096.webp)
96
Agilent 1260 Infinity VWD User Manual
7
Error Information
Detector Error Messages
No heater current
Error ID: 7453
The lamp heater current in the detector is missing. During lamp ignition,
the processor monitors the heater current. If the current does not rise
above the lower limit within 1 , the error message is generated.
Probable cause
Suggested actions
1
Lamp disconnected.
Ensure the lamp is connected.
2
Ignition started without the top foam in
place.
Please contact your Agilent service
representative.
3
Fan not running (permitting lamp on).
Please contact your Agilent service
representative.
4
Defective main board.
Please contact your Agilent service
representative.
5
Defective or non-Agilent lamp.
Exchange the lamp.
6
Defective power supply.
Please contact your Agilent service
representative.
Содержание G1314B
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Страница 30: ...30 Agilent 1260 Infinity VWD User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Страница 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Страница 190: ...190 Agilent 1260 Infinity VWD User Manual 12 Hardware Information Early Maintenance Feedback EMF...
Страница 207: ......