Agilent 1260 Infinity VWD User Manual
3
Installing the Detector
Flow Connections to the Detector
Flow Connections to the Detector
Tools required
Description
Wrench, 1/4 – 5/16 inch
(for capillary connections)
Parts required
#
p/n
Description
1
G1314-68755
Accessory kit
Hardware required
Other modules depend on system setup
Preparations
Detector is installed in the LC system.
WA R N I N G
Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
➔
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow good
laboratory practice.
➔
The volume of substances should be reduced to the minimum required for the
analysis.
➔
Do not operate the instrument in an explosive atmosphere.
N O T E
The flow cell is shipped with a filling of isopropanol (also recommended when the
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
Содержание G1314B
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Страница 30: ...30 Agilent 1260 Infinity VWD User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Страница 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Страница 190: ...190 Agilent 1260 Infinity VWD User Manual 12 Hardware Information Early Maintenance Feedback EMF...
Страница 207: ......