Agilent 1260 Infinity VWD User Manual
12
Hardware Information
Firmware Description
Firmware Description
The firmware of the instrument consists of two independent sections:
•
a non- instrument specific section, called
resident system
•
an instrument specific section, called
main system
Resident System
This resident section of the firmware is identical for all Agilent
1100/1200/1220/1260/1290 series modules. Its properties are:
•
the complete communication capabilities (CAN, LAN and RS- 232C)
•
memory management
•
ability to update the firmware of the 'main system'
Main System
Its properties are:
•
the complete communication capabilities (CAN, LAN and RS- 232C)
•
memory management
•
ability to update the firmware of the 'resident system'
In addition the main system comprises the instrument functions that are
divided into common functions like
•
run synchronization through APG remote,
•
error handling,
•
diagnostic functions,
•
or module specific functions like
•
internal events such as lamp control, filter movements,
•
raw data collection and conversion to absorbance.
Firmware Updates
Firmware updates can be done using your user interface:
•
PC and Firmware Update Tool with local files on the hard disk
Содержание G1314B
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Страница 30: ...30 Agilent 1260 Infinity VWD User Manual 2 Site Requirements and Specifications Performance Specifications...
Страница 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Страница 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Страница 190: ...190 Agilent 1260 Infinity VWD User Manual 12 Hardware Information Early Maintenance Feedback EMF...
Страница 207: ......