Agilent 1260 Infinity VWD User Manual
13
Introduction to the Variable Wavelength Detector
1
Optical System Overview
G1314B/C Lamp
The light source for the UV wavelength range is a deuterium lamp. As a
result of plasma discharge in a low pressure deuterium gas, the lamp
emits light over the 190 to 600 nm wavelength range.
Source Lens Assembly
The source lens receives the light from the deuterium lamp and focuses it
onto the entrance slit.
Entrance Slit Assembly
The entrance slit assembly has an exchangeable slit. The standard one has
a 1- mm slit. For replacement and calibration purposes to optimize the
alignment, a slit with a hole is needed.
Filter Assembly
The filter assembly is electromechanically actuated. During wavelength
calibrations it moves into the light path.
The filter assembly has two filters installed and is processor- controlled.
A photo sensor determines the correct position.
OPEN
nothing in light path
CUTOFF
cut off filter in light path at
λ
> 370 nm
HOLMIUM
holmium oxide filter for wavelength check