Agilent 1260 Infinity VWD User Manual
121
Maintenance
9
Overview of Maintenance
Overview of Maintenance
The following pages describe maintenance (simple repairs) of the detector
that can be carried out without opening the main cover.
Table 13
Simple Repairs
Procedures
Typical Frequency
Notes
Deuterium lamp
exchange
If noise and/or drift exceeds your application limits or
lamp does not ignite.
A VWD test should be performed after
replacement.
Flow cell exchange
If application requires a different flow cell type.
A VWD test should be performed after
replacement.
Cleaning flow cell
parts cleaning or
exchange
If leaking or if intensity drops due to contaminated flow
cell windows.
A pressure tightness test should be
done after repair.
Leak sensor drying
If leak has occurred.
Check for leaks.
Leak handling system
replacement
If broken or corroded.
Check for leaks.