52
Troubleshooting and Test Functions
Power Supply Indicator
Power Supply Indicator
The power supply indicator is integrated into the main power switch. When
the indicator is illuminated (
green
) the power is ON.
Instrument Status Indicator
The instrument status indicator indicates one of four possible instrument
conditions:
•
When the status indicator is
OFF
(and power switch light is on), the
instrument is in a
prerun
condition, and is ready to begin an analysis.
•
A
green
status indicator, indicates the instrument is performing an analysis
(
run
mode).
•
A
yellow
indicator indicates a
not-ready
condition. The instrument is in a
not-ready state when it is waiting for a specific condition to be reached or
completed (for example, front cover not installed), or while a self-test
procedure is running.
•
An
error
condition is indicated when the status indicator is
red
. An error
condition indicates the instrument has detected an internal problem which
affects correct operation of the instrument. Usually, an error condition
requires attention (for example, leak, defective internal components). An
error condition always interrupts the analysis.
Содержание G1313A
Страница 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Страница 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 12: ...12 Contents ...
Страница 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Страница 49: ...3 3 Troubleshooting and Test Functions The modules built in troubleshooting and test functions ...
Страница 93: ...93 Troubleshooting and Test Functions Step Commands ...
Страница 159: ...5 5 Parts and Materials Detailed illustrations and lists for identification of parts and materials ...
Страница 160: ...160 Parts and Materials Main Assemblies Main Assemblies Figure 19 Autosampler Main Assemblies 5 3 4 2 1 9 8 7 6 ...
Страница 166: ...166 Parts and Materials Analytical Head Assembly Figure 21 Analytical Head Assembly 2 3 4 5 6 1 ...
Страница 191: ...191 Parts and Materials Remote Cables NOTE START and STOP are connected via diodes to pin 3 of the 3394 connector ...
Страница 216: ...216 Introduction to the Autosampler Electrical Connections ...
Страница 217: ...7 7 Theory of Operation Theory of operation of mechanical hardware electronics and instrument interfaces ...
Страница 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Страница 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Страница 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Страница 269: ...9 9 Specifications Performance specifications of the Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Страница 274: ...274 Specifications Performance Specifications ...
Страница 275: ...10 Legal Safety and Warranty Information ...
Страница 286: ...286 Agilent Technologies on Internet ...
Страница 290: ...290 Index ...
Страница 291: ......